Keep Vein Open
Alarm will sound when pump goes into Keep Vein Open mode.
Press enter button (4) to silence the alarm
Keep Vein Open will only function if the bag volume is greater than the VTBI
Pump Unattended
Intermittent alarm will sound when the pump is on hold or no button
has been pressed or pump started. This will last for 1 hour.
Press enter button (4) confirm settings and start infusion
Alarms
When there is an immediate problem the pump will Alarm and stop functioning until
the problem has been resolved.
Downstream Occlusion
Indicates there is a problem between the pump and the patient.
Press enter button (4) to silence the alarm
Rectify the problem and pump will automatically restart
Upstream Occlusion or Air in Line
Indicates there is a problem between the bag and the pump.
Press enter button (4) to silence the alarm; rectify the problem then restart
by pressing stop/start button twice (5)
Depleted Internal Battery
Indicates the internal battery requires renewing.
Press enter button (4) to silence the alarm.
Turn the pump off using the on/off button (2)
and replace the internal battery
Turn the pump on using the on/off button (2) and restart pump by selecting
resume after power off programme and pressing stop/start button twice (5)