128957_V01_QSG_ASSIST_PLUS_EN.docx 3/4
Setting up pipetting programs
You can easily create multi-step programs with the VIALAB software on a PC or select a predefined
program on the pipette and adapt the parameters to your needs.
Contains all programs created with VIALAB software.
Aspirate a transfer volume followed by a mix.
Dispense multiple aliquots of the same volume.
Dispense multiple aliquots of different volumes.
Aspirate multiple aliquots of different volumes.
Transfer samples between plates of the same well number.
Reformat 12, 24, 48, 96 and 384 well plates.
Create multi-stepped user-defined pipetting programs.
The Toolbox provides a variety of options, like pairing ASSIST PLUS with a pipette, device
information, tip monitoring, tip type, park position and more.
Always switch off the device and disconnected form the electricity supply when carrying
out maintenance work.
Clean the ASSIST PLUS housing a lint-free cloth lightly soaked with mild soap solution in
distilled water or with a 70 % solution of isopropanol or ethanol.
To decontaminate the ASSIST PLUS, take out the Waste bin (h) to the front, remove the waste
bin frame, the magnetically attached Sensor reflector (j) and the decks. Wipe the clean
instrument surface with a lint-free cloth lightly soaked with a suitable disinfectant and wipe dry
directly.
Decontaminate the Sensor reflector by spraying and wiping it with 70% Ethanol and after a
short treatment time (1-2 minutes) wiping it dry.
The aluminum decks and the waste bin with frame can be autoclaved at 121 °C for 20 min.
ASSIST PLUS must not be disposed of with unsorted municipal waste.
Dispose of ASSIST PLUS in accordance with the regulations in your area governing
disposal of devices.