EasyManua.ls Logo

Integra Codman DirectLink User Manual

Integra Codman DirectLink
7 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
For complete operating instructions, please refer to the manufacturer’s user manual
DirectLink® ICP Module
For the Direct Connection of the
Codman Microsensor
®
ICP Transducer
to a Patient Monitoring System
Reference Signal
Indicators
Status Lights
Arrow Buon
(Patient Monitor
Calibration)
Zero Buon
(ICP Sensor Zeroing)
Output Connection
(To Patient Monitor)
Input Connection
(To Extension Cable & ICP Sensor)
A DIVISION OF INTEGRA LIFESCIENCES integralife.com
QUICK REFERENCE GUIDE
Question and Answer IconNeed help?

Do you have a question about the Integra Codman DirectLink and is the answer not in the manual?

Integra Codman DirectLink Specifications

General IconGeneral
Product LineCodman
ModelDirectLink
CategoryControl Unit
ManufacturerIntegra LifeSciences
CompatibilityCodman CERTAS Plus Programmable Valves, Codman HAKIM Programmable Valves
FunctionWireless control and programming of programmable valves.
InterfaceWireless
Power SourceBattery
Product CategoryMedical Device
Intended UseThe Codman DirectLink System is intended to provide wireless control of Codman CERTAS Plus Programmable Valves and Codman HAKIM Programmable Valves.

Summary

Patient Monitor Setup

Powering the ICP Module & Calibrating to the Patient Monitor

Connect the ICP Module to the patient monitor, wait for solid amber status light, then follow calibration steps using the arrow button.

ICP Sensor Setup

Zero the ICP Sensor Prior to Implantation

Connect ICP Extension Cable and Transducer to the module. Verify amber status, press and hold ZERO BUTTON until green light appears.

Disconnecting & Reconnecting

Patient Monitor for Transport and Patient Care Activities

Reconnect module and sensor to the same module used for zeroing. Press ARROW and ZERO buttons to confirm zero and continue monitoring.

Troubleshooting

SOLID RED LIGHT – MODULE ERROR DETECTED

Explains the reset procedure for a solid red light (disconnect/reconnect power) and troubleshooting steps for self-test and during use errors.

FLASHING AMBER LIGHT WILL NOT TURN GREEN

Discusses issues with non-stable signals during zeroing, sensor implantation, and resuming monitoring with the acknowledge function.