Do you have a question about the Integra Codman DirectLink and is the answer not in the manual?
Product Line | Codman |
---|---|
Model | DirectLink |
Category | Control Unit |
Manufacturer | Integra LifeSciences |
Compatibility | Codman CERTAS Plus Programmable Valves, Codman HAKIM Programmable Valves |
Function | Wireless control and programming of programmable valves. |
Interface | Wireless |
Power Source | Battery |
Product Category | Medical Device |
Intended Use | The Codman DirectLink System is intended to provide wireless control of Codman CERTAS Plus Programmable Valves and Codman HAKIM Programmable Valves. |
Connect the ICP Module to the patient monitor, wait for solid amber status light, then follow calibration steps using the arrow button.
Connect ICP Extension Cable and Transducer to the module. Verify amber status, press and hold ZERO BUTTON until green light appears.
Reconnect module and sensor to the same module used for zeroing. Press ARROW and ZERO buttons to confirm zero and continue monitoring.
Explains the reset procedure for a solid red light (disconnect/reconnect power) and troubleshooting steps for self-test and during use errors.
Discusses issues with non-stable signals during zeroing, sensor implantation, and resuming monitoring with the acknowledge function.