EasyManua.ls Logo

Interacoustics AS608 - Operating and Environmental Conditions; Operating Temperature Limits; Heat Source Avoidance

Interacoustics AS608
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AS608/AS608e Operation Manual Page27
Operating Temperature
Keep away from heat:
Do not site the AS608/AS608e next to a radiator or any
other heat source. Allow six inches between the back panel
and the wall to ensure proper ventilation.
Operating temperature: 15C- 35C/ 59-95F:
In operation the instrument should not be subject to
temperatures below 15C or above 35C/ below 59F or
above 95F.
If it has been exposed to abnormal low temperatures the
warm-up time must be respectively prolonged.
Care and Maintenance
The performance and reliability of the AS608/AS608e will be
prolonged if the following recommendations for care and
maintenance are adhered to:
Using an external power Supply e.g. ASA30M-0301 - Turn the
switch to the off position.
Great care when handling the headset:
Great care should be exercised when handling the headset as
dropping it may alter the calibration.
Annual calibration headset:
The AS608/AS608e has been designed to provide many
years of reliable service, but annual calibration is
recommended.
We do also recommend to calibrate the AS608/AS608e if
something drastic happens to part of it (e.g. headset is
dropped).

Other manuals for Interacoustics AS608

Related product manuals