ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuturereference.

Questions and Answers

  • L
    Lisa SmithJul 31, 2025
    What to do if O below normal (YELLOW) light stays on Invacare Air Compressor?
    • S
      Samantha RobertsonJul 31, 2025
      First, make sure the concentrator has warmed up for at least 20 minutes. If it hasn't, turn the compressor off until it warms up. If the yellow light is still on, inspect the connection between the compressor and the concentrator for any damage or pinching. If damaged, replace it. If pinched, straighten the tubing. Turn the compressor off for at least 10 seconds, then turn it back on. If the yellow light remains on after 10 minutes, the concentrator may need service.
  • M
    Marc JohnsonAug 5, 2025
    What to do if the green filling light stays on for Invacare Air Compressor?
    • K
      Kevin Mason DDSAug 5, 2025
      First, ensure the cylinder regulator is set to Off. Next, check the recommended fill times in the cylinder’s manual. Then, make sure the cylinder is properly connected by turning the compressor off, removing and reinstalling the cylinder, and turning the compressor back on. Also, inspect the compressor/concentrator line for proper connection and any damage like leaks or kinks. Finally, verify the concentrator liter flow setting is at three L/min or less, adjusting if needed. If the green filling light remains on, the compressor or concentrator may need service.
  • M
    Mrs. Tracy Ward MDAug 9, 2025
    What to do if red light is on Invacare Air Compressor?
    • L
      luis16Aug 9, 2025
      First, check the cylinder gauge to see if it is full. If the cylinder is full, remove it and turn the unit off. Then, perform the Cylinder Prefill Inspection with a cylinder that is not full. If the cylinder is not full, ensure that the cylinder is connected properly. Turn the compressor off, remove the cylinder, and reinstall it. Then, turn the compressor back on. If the cylinder is not full and is connected properly, the compressor may have an internal failure.
  • J
    Jacob WilliamsAug 13, 2025
    Why is my Invacare HomeFill Oxygen System cycling on and off?
    • M
      Michael RoweAug 13, 2025
      Check that the concentrator liter flow setting is at 3 L/min or less. If it’s higher, adjust the liter flow setting and wait at least 20 minutes for the green light to appear. If the green light does not appear after adjusting the flow, the concentrator may need service.
  • J
    James OlsenAug 18, 2025
    How to fix Invacare HomeFill Oxygen System Air Compressor with no indicator lights?
    • G
      ggarciaAug 18, 2025
      Turn the compressor off. Check that the power cord is properly plugged into both the compressor and the wall outlet. Turn the compressor on again. Clean the air filters. If the indicator lights still don't come on after these steps, the unit may need service.
  • L
    Lisa AndersonAug 22, 2025
    Why is my Invacare HomeFill Oxygen System regulator hissing when off?
    • B
      Betty NelsonAug 22, 2025
      Turn the flow selector off. If the hissing sound continues, the unit may need service.
  • Q
    qgutierrezAug 27, 2025
    What to do if oxygen is not being delivered with Invacare HomeFill Oxygen System?
    • E
      Elizabeth DavisAug 27, 2025
      Refill the cylinder. Check that the cannula is straight and not pinched; replace it if damaged. If the oxygen is still not being delivered, the unit may need service.

Summary

General Information

Intended Use and Critical Warnings

Defines the product's intended use and highlights critical safety warnings.

Safety Precautions

General Safety Guidelines

Provides essential safety instructions, warnings, and precautions for product operation.

Product Setup

Unpacking the Compressor

Instructions for unpacking the compressor and its components.

Initial Setup Checklist

Checklist for users to complete before operating the compressor for the first time.

Operating the System

Cylinder Prefill Inspection

Procedure for inspecting cylinders before refilling to ensure safety and proper function.

Connecting Cylinder to Compressor

Step-by-step instructions for securely connecting the oxygen cylinder to the compressor.

Troubleshooting Guide

Problem-Solution Guide

Lists common operational problems and provides step-by-step solutions.

Product Warranty

Invacare HomeFill Oxygen System Specifications

General IconGeneral
BrandInvacare
ModelHomeFill Oxygen System
CategoryAir Compressor
LanguageEnglish

Related product manuals