What to do if O below normal (YELLOW) light stays on Invacare Air Compressor?
- SSamantha RobertsonJul 31, 2025
First, make sure the concentrator has warmed up for at least 20 minutes. If it hasn't, turn the compressor off until it warms up. If the yellow light is still on, inspect the connection between the compressor and the concentrator for any damage or pinching. If damaged, replace it. If pinched, straighten the tubing. Turn the compressor off for at least 10 seconds, then turn it back on. If the yellow light remains on after 10 minutes, the concentrator may need service.