SECTION 3—CLEANING AND DISINFECTING
Stratos™Compact 22 Part No 1130203
SECTION 3—CLEANING AND
DISINFECTING
Following the cleaning instructions after each use will prevent
any remaining medication in the bottle from drying and
causing the device to nebulize ineffectively.
It is recommended that the disposable nebulizer and the
mouthpiece be thoroughly rinsed with hot tap water after
each use and cleaned with a mild detergent after the last
treatment of the day. The same pieces MUST also be
disinfected daily by soaking in a vinegar/water solution or a
commercially available disinfectant. If your physician or
respiratory therapist specifies a different cleaning procedure,
follow their instructions
Cleaning the Outer Case
WARNING
Unplug unit before cleaning.
DO NOT submerge in water to clean. Clean outer case
with damp cloth.