SECTION 4—MAINTENANCE
Stratos™Compact 28 Part No 1130203
5. Store the device in a safe area.
NOTE: Store the repackaged device in a dry area. DO NOT place
other objects on top of the repackaged device.
Nebulizer Change
The nebulizer MUST be replaced after a long period of
inactivity, or when it is obstructed by dry medication, dust,
etc. The reusable nebulizer should last from six months to one
year with general use. The disposable nebulizer should last 30
days.
Filter Change
1. The filter MUST be replaced after approximately 30 hours
of use or when it turns grey.
2. Open the filter holder to replace the filter with a new one.
CAUTION
DO NOT use cotton or any other material.
DO NOT wash or clean the filter.
Use only Invacare filters.
DO NOT operate unit without an air filter.