SECTION 3—CLEANING AND DISINFECTING
Stratos™Compact 24 Part No 1130203
2. Disassemble nebulizer chamber by turning
counterclockwise and separating.
3. Fill two plastic containers or bowls:
A. Washing: Hot water/detergent solution.
B. Soaking: Hot water/vinegar solution (one part vinegar
to three parts water).
4. Thoroughly clean the nebulizer and mouthpiece in hot
water/detergent solution.
5. Remove from the water/detergent solution and rinse with
clear hot tap water.
6. Soak in hot water/vinegar solution for 30 minutes.
7. Rinse with hot tap water again and air dry thoroughly.
NOTE: If using medical disinfectant cleaners, follow
manufacturer’s instructions carefully.
8. Keep the outer surface of the tubing dust-free by wiping
regularly.
NOTE: The nebulizer air tubing does not have to be washed
internally because only filtered air passes through the tubing.