Service Information
31 Canis014D07 User Manual
6 Service Information
6.1 Service Life
The estimated product lifetime is up to 7 years under appropriate regular inspection and maintenance.
The product life is decided by that of the equipment.
For other replaceable parts, their service life will not affect the life cycle of the whole product.
Main parts (parts required to maintain the function of the product) of this product will be
stocked for 5 years after discontinuance of production for repairing.
6.2 Periodic Inspection and Maintenance
To ensure image quality and the safety for the patient and operator, it is highly recommended that periodic inspection
and maintenance be carried out by the qualified maintenance personnel.
Before operation, check if the equipment is clean and ready for
use
After each use, check if the equipment has been turned off
Check if the screen can be displayed normally when starting.
Check if the exposure indicator light turns on when the exposure
button is pressed.
Check if the buzzer sound works OK
Check if the battery charging LED indicator comes on when
charging the battery.
Check if all visible labels are intact and legible.
Check the battery life and check whether the battery is swollen
The enssential performance of the equipment should be tested by
the local third-party organization.