EasyManua.ls Logo

iRay Technology Mars1417X - Page 72

iRay Technology Mars1417X
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Wireless Digital Wireless Digital Flat Panel Detector Mars1417X
Group 1, Class B
Harmonic
distortion
IEC 61000-3-2
Class A
Professional healthcare facility environment
Voltage
fluctuations and
flicker
IEC 61000-3-3
Compliance
Professional healthcare facility environment
6.2.2. EMS Compliance Table
Enclosure Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM
field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Near fields from
RF wireless
communications
equipment
IEC 61000-4-3
Refer to table “Near fields from RF wireless
communications equipment”
Rated power
frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
Near fields from RF wireless communications equipment
Test
frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
1700-1990
Pulse modulation 217Hz, 28V/m
1845
1970
62 User Manual of Mars1417X