6.Regulatory Information
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
Input a.c. power port
Phenomenon Basic EMC standard
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Line-to-line
IEC 61000-4-5 ±0.5 kV, ±1 kV
Line-to-ground
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
Voltage dips
IEC 61000-4-
11
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and
315º
and
70% UT; 25/30 cycles
Single phase: at 0º
interruptions
11
0% UT; 250/300 cycles
Recommended separation distances between portable or mobile RF
communication device and detector:
Portable RF communications equipment, including antennas, can effect medical electrical
equipment. The warning should include a use distance such as “be used no closer than 30 cm (12
inches) to any part of the [ME EQUIPMENT or ME SYSTEM], including cables specified by the
manufacturer”.
Cable provided for EMC
Cable
Recommended
length
Shielded/Unshielded Number
Cable
classification
AC power cable 1.8m Unshielded 1 piece AC power
Electromagnetic Compatibility (EMC)
Mars1417X requires special precautions regarding EMC and needs to be installed only by iRay
or authorized personnel and put into service according to EMC information provided in the user
manual. Mars1417X in use may be susceptible to electromagnetic interference from portable and
mobile RF communications such as mobile (cellular) telephones. Electromagnetic interference may
result in incorrect operation of the system and create a potentially unsafe situation.
Mars1417X conforms to this EN60601-1-2:2015 standard for both immunity and emissions.
71 User Manual of Mars1717X