TTI-7 Iss.02 06/11
The TRIGger:COUNt? <slx> command queries the specified scanning list timer cycle count.
<slx> SL1, SL2, SL3, SL4 Specify scan list
Returns the cycle count value (1 to 9999), or “CONT”, if the value is continuous.
TRIGger:DELay <hh:mm:ss>
TRIGger:DELay <slx>, <hh:mm:ss>
The TRIGger:DELay <hh:mm:ss> command is used to set the general timer cycle delay time.
The TRIGger:DELay <slx>, <hh:mm:ss> command is used to set the individual scanning list timer cycle delay time.
<slx> SL1, SL2, SL3, SL4 Specify scan list
<hh:mm:ss> 00:00:00 to 99:59:59 Specify timer delay
Terminates the current measurement cycle and scanning routine, and sets the trigger mode to HOLD.
TRIGger:DELay?
TRIGger:DELay? <slx>
The TRIGger:DELay? Command queries the general timer cycle delay time.
The TRIGger:DELay? <slx> command queries the specified scanning list timer cycle delay time.
<slx> SL1, SL2, SL3, SL4 Specify scan list
Returns the delay time (00:00:00 to 99:59:59).
SAMPle Group
SAMPle:COUNt?
SAMPle:COUNt? <slx>
The SAMPle:COUNt? command queries the general timer sample count. This is the same as the general timer cycle
count.
The SAMPle:COUNt? <slx> command queries the specified scanning list sample count. The sample count is the
product of the timer cycle count and the number of active channels in the specified scanning list.
<slx> SL1, SL2, SL3, SL4 Specify scan list
Returns the sample count (1 to 9999), or “CONT”, if the value is continuous.
SAMPle:DELay <slx>, <hh:mm:ss>
This command is used to set the specified scanning list timer sample delay.
<slx> SL1, SL2, SL3, SL4 Specify scan list
<hh:mm:ss> 00:00:00 to 99:59:59 Specify trigger delay
DEFault Trigger delay 00:00:00
Changing a timer sample delay for a scanning list that is currently in use will result in the instrument terminating the
current measurement cycle and scanning routine. Also, the Run/Hold trigger mode will be set to HOLD.