Emissions test Compliance Electromagnetic environment — guidance
RF emissions
CISPR 11
Group 1
This unit uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
This unit is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
This unit is intended for use in the electromagnetic environment specified below.
The customer or the user of this unit should assure that it is used in such an environment.
Guidance and manufacturer’s declaration
–
electromagnetic emissions
• Medical electronic devices are designed to ensure electromagnetic compatibility (EMC).
• These devices must be installed and used in accordance with the EMC information provided in
the attached document.
• Do not use portable and/or mobile RF communication devices closer than 30 cm within the
medical electronic device. If it is brought closer than 30 cm, the performance of the medical
equipment may deteriorate.
• Electrode Cord length 2 m
• If accessories other than those supplied as spare parts by the manufacturer are used, the
emission of this unit may increase and immunity may be reduced.
• Do not place this unit next to or on top of another device when using it. If it has to be placed
next to or on top of another device, check that this unit and the device function properly before
use.
EMC