16
16
Select the output mode using the Output-
mode selector key.
Turn on the unit.
If all 6 Output intensity knobs are not set to “0,”
a warning sound will be emitted (4 intermittent
beeps) for 15 seconds and the Error lamp will
come on when the unit is turned on.
If the output level is set to H (High), the buzzer
will sound (2 intermittent beeps) for 15 seconds.
Press the Buzzer key to mute the beeping.
If the output level is switched from the L level to the H level with the unit turned
on, a warning sound will be emitted for 15 seconds. the Error lamp will come on,
and the output will be cut off. In such a case, put all the Output intensity knobs
back in the “0” position. Otherwise, it will not be possible to start the treatment.
Incidentally, when the output is set to the H level, the output-H warning sound (2
intermittent beeps) will be emitted continuously. The warning sound can be
muted by pressing any one of the Buzzer key, Battery check key, Output-mode
selector key, Frequency key, Phase duration key, and Timer key.
When the unit is turned on with the Output level selector switch set to H, the
warning sound will be emitted (2 intermittent beeps).
H (High)-level output : Max. 32 mA (peak)
L (Low)-level output : Max. 16 mA (peak)
4
5
H level
Error lamp
Output-mode selector key
L (low) output level must be used in
Electroacupuncture mode