DIGISPENSE 3009 CONTROLLER MODULE, STYLE B Page 3-87
29 October, 2019 IVEK Corp. P03R017E
Table 3.6 Output 1 – Output 4, AUX Out 1 and Aux Out 2 Signal Description
(18 characters maximum)
Operating States
STATUS (indicated in upper-right-hand
corner of LCD)
<DISABLED>
<BUSY> DISCHARGE
<BUSY> AP PRIME
<BUSY> PRD OP
<BUSY> FLD OP
<BUSY> PRD+FLD OPS
<BUSY> ALL OPS
<READY> IDLE
<READY> PRODUCTION
<SUCCESS> LAST
<SUCCESS> PULSE
<STATUS> FAULT
<STATUS> REFERENCE
<STATUS> RECIPE
<LINK> WITH INPUT
No Operation
INITIALIZING (power-up)
IDLE
T
T
T
3
IDLE (production mode disabled)
T
T
3
IDLE (fluid mode disabled)
T
T
T
3
NOT CONFIGURED
T
3
MOTION LOCKED
T
3
AUTOTRIG IDLE
T
3
FAULTED
T
CLEARING FAULT
T
Reference Operation
REFERENCING
T
T
REFERENCE REQUIRED (DS3009 recommended)
T
T
2
T
3
T
Fluidic Setup Mode Operation
PRIMING
T
1
T
T
T
Agitate Operation
ISOLATING
T
T
T
AGITATING
T
T
T
AGITATE DWELL
T
T
T
RETURNING
T
T
T
Production Mode Operation
PRE-OP DWELL
T
T
T
DISPENSING or METERING
T
T
T
T
DRAWBACK DWELL
T
T
T
DRAWBACK
T
T
T
POST-OP DWELL
T
T
T
T
Setup E Screen (D = default setting)
OUTPUT AUX 1 TRUE = CONDUCTING
D
OUTPUT AUX 2 TRUE = CONDUCTING
D
OUTPUT 1 TRUE = CONDUCTING
D
OUTPUT 2 TRUE = NOT CONDUCTING
D
OUTPUT 3 TRUE = CONDUCTING
D
OUTPUT 4 TRUE = CONDUCTING
D
Notes
T = True, A blank indicates "F", False
T
1
- AP Prime will not go "TRUE" until 2nd rotation of prime operation
T
2
- on DS3009 only, will show "TRUE" if a reference is recommended
T
3
- Only "T" if last production operation completed without fault.
Each output is configurable for TRUE = CONDUCTING or NOT CONDUCTING