DIGISPENSE 3009 CONTROLLER MODULE, STYLE B Page 3-98
29 October, 2019 IVEK Corp. P03R017E
y2 y2,<value2> Returns the current Motor/Base selection.
y2,<value2> y2,<value2> Sets the Motor/Base configuration. (Refer to Table 3.2)
<value2>:
Maximum: 7
Minimum: 0 (default)
CAUTION
The Motor/Base Module number selection in the Setup C screen MUST match the Motor/Base Module
attached to the Controller Module. An incorrect setting could overheat the motor and damage the equipment
and cause a hazardous condition.
y3 y3,<value2> Returns the current Stop Position.
y3,<value2> y3,<value2> Sets the stop position in degrees.
<value2>:
Maximum: 330
Minimum: 0
Default: 90
y4 y4,<value2> Returns the current Meter Stop mode.
y4,<value2> y4,<value2> Sets the meter stop mode.
<value2>:
0 = Stop Position (default)
1 = Immediate
y5 y5,<value2> Returns the current Stalls per Fault.
y5,<value2> y5,<value2> Sets the stalls per fault.
<value2>:
Maximum: 15
Minimum: 1
Default: 4
y11 y11,<value2> Returns the current Initial Rate selection.
y11,<value2> y11,<value2> Sets the Initial Rate. (Refer to Table 3.2)
<value2>:
0 = Low
1 = Medium
2 = High (default)
z => SOFTWARE VERSION
z or z0 z0,<value2> Returns the software version as text.
z1 z1,<value2> Returns the Current device type (DS3009)
z2 z2,<value2> Returns the firmware CRC
3.4.10 Warnings
Warnings indicate problems in the command received, or a state of the Motor/Base Module which prohibits immediate
operation. An appropriate command (other than ‘clear faults’) may be required to operate the pump.
1 Command Not Valid - Response to any unrecognized command.
2 Value Not Valid - Response to any out of range value.
5 Production Mode is Disabled - An attempt was made to run in production mode while it is disabled.
8 Serial In Motion Not Enabled - Serial command sent to begin motion while motion disabled due to serial
command. (k0)
11 Second Command Character - A second command character (alphabetic character) was seen in a single
command (before <CR> character). Entire command is ignored.
15 Descriptor Not Valid - First numerical value not valid for command letter.
16 Recipe Is Blank
17 Recipe Is Not Blank
18 Motion Is Disable Via Logic - Serial command sent to begin motion while motion disabled due to logic input.