MANUAL DE INSTRUCCIONES CODIGO 80063 REV E 24/12/09 (It can be modified without notice) Pag.: 11
J.P. SELECTA s.a. Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España Tel (34) 93 770 08 77 Fax (34) 93 770 23 62
e-mail: selecta@jpselecta.es - http://www.jpselecta.es
CALIBRATE:
Ajuste de calibración del aparato
TERMINAL ID:
Número del terminal de puerto serie.
ALARMAS:
OVERTEMP:
Exceso de temperatura
Función:
Desactiva estado de run.
Activa alarma correspondiente.
Desactiva la calefacción.
Para el motor.
Desactivación:
Para desactivar la alarma pulsaremos P5
(START/ STOP) y pasaremos al estado de STOP.
UNDERTEMP:
Nivel bajo de temperatura
Función:
Desactiva estado de run.
Activa alarma correspondiente.
Desactiva la calefacción.
Para el motor.
Desactivació:
Para desactivar la alarma pulsaremos P5
(START/ STOP) y pasaremos al estado de STOP.
LEVEL:
Termostato de seguridad o de nivel
Función:
Desactiva estado de run.
Activa alarma correspondiente.
Desactiva la calefacción.
Para el motor.
Desactivación:
Para desactivar la alarma pulsaremos P5
(START/ STOP) y pasaremos al estado de STOP.
100.0ºC 100u 99h59m
STAT: CFG CALI +10.0
100.0ºC 100u 99h59m
STAT: ALARM OVERT.
100.0ºC 100u 99h59m
STAT: ALARM UNDERT.
100.0ºC 100u 99h59m
STAT:ALARM TTO SAFETY
100.0ºC 100u 99h59m
STAT: CFG TERM ID: 01
CALIBRATE:
Calibration adjustment of the bath.
TERMINAL ID:
Number of the serial port terminal.
ALARMS:
OVERTEMP:
Temperature too high.
Function:
Desactivate run mode.
Start the alarm.
Stop the heating.
Stop the motor.
Solution:
To stop the alarm press P5 (START/
STOP) and go to STOP mode.
UNDERTEMP:
Temperature too low.
Function:
Desactivate run mode.
Start the alarm.
Stop the heating.
Stop the motor.
Solution:
To stop the alarm press P5 (START/
STOP) and go to STOP mode.
LEVEL:
Safety or level thermostat.
Function:
Desactivate run mode.
Start the alarm.
Stop the heating.
Stop the motor.
Solution:
To stop the alarm press P5 (START/
STOP) and go to STOP mode.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the J.P. SELECTA UNITRONIC-ORBITAL and is the answer not in the manual?

J.P. SELECTA UNITRONIC-ORBITAL Specifications

General IconGeneral
BrandJ.P. SELECTA
ModelUNITRONIC-ORBITAL
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals