JIF-2002 WEIGHING INDICATOR&CONTROLLER
31
2. Select Digital Calibration
An easy way to make calibration by inputting Load Cell’s Full Scale Output
voltage
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.
STEP 3: Please press GROSS/NET key and a blinking CAL will show
on the screen. Then press the key. The screen will show
F-CAL. Please use the or key to choose digital Calibration
(d-CAL).
STEP 4: The screen will show F-CAL. Please use the or key to
choose digital Calibration (d-CAL). Please press the key.
1Setting Minimum Division
The display of di 1 shows the smallest division. Use
theor key to move through the available divisions.12
5102050. Press the key to set the minimum division.
2Setting Decimal
F000 will change---see 4-2
The screen will show dp d000.000. A blinking decimal will
show on the screen. Use theor key to move through the
available decimal
Point
position. Press the key to set
the decimal position.
3Setting Maximum Capacity
When setting maximum capacity, the screen will show CAP
C000000.Use theor key to move through digits. Use