JIF-2002 WEIGHING INDICATOR&CONTROLLER
32
the or key to set the numeric value. Press the ↵ key to
finish the step.
﹝4﹞Setting Full Scale Output Voltage of the Load Cell Sensors
When setting full scale output voltage of the load cell sensors,
the screen will show LC-CAP→L000000.. Please use the
or key to move through digits. Use the or key to set
the numeric value. Press the ↵ key to finish the step.
﹝5﹞ZERO Adjust
The Screen will display ZERO . Please move the calibration
mass and objects away on the Weighing device then press ↵
key. A display of . . . . . . . means finishing the Adjustment.
﹝6﹞d-SPAN Calibration
The screen will show d-SPAN . Press ↵ key and place your
calibration mass on the weighing device and input weight value.
Use the or key and the or key to enter Load Cell O/P
Volt. Please press the ↵ key to finish d-SPAN. The screen will
show . . . . . . . .
STEP 5: The screen will show END .
Slide the SET switch to the original side.
※Example of selecting FULL CALIBRATION
(Div 2, 3 decimal, Max cap.20)