JIF-2002 WEIGHING INDICATOR&CONTROLLER
36
4-7 CLEAR SET POINT DATA
STEP 1: Turn the Power Switch OFF on the rear panel. Slide the SET
switch to the OFF side (The SET switch remain in OFF status).
STEP 2: Turn the power switch ON. The screen will show Normal
operation condition.
STEP 3: Please press STANDBY key and hold SET POINT/ CODE key
at the same time until the screen shows C Lr cd .Please
release the STANDBY key ( SET POINT/CODE key still
holding). Please release the SET POINT/CODE key in
sequence. Please press the key and the JIF-2002 will
subsequently ask the operator to clear set point data.
STEP 4: Use the or key to choose NO or YES and press the key to
confirm.
4-8 SYSTEM INITILAIZE
Re-install resets the JIF-2002 to the initial factory settings. Use
Re-install only if you want to return Function, Set Point or Calibration
to their initial settings.
STEP 1: Turn the Power Switch OFF on the Rear Panel, and slide SET
switch to he set side.
STEP 2: Turn the power switch ON. The screen will show blinking
SELECT.