EasyManua.ls Logo

JFE Advantech RINKO FT - Page 42

Default Icon
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Calibration coefficients:
Coefficients
C0=F
C1=F
C2=F
d0=F
d1=F
d2=F
d3=F
d4=F
Cp=F
e0=F
FilmNo=
S
docaldate=S
A=F
B=F
C=F
D=F
E=F
F=F
G=F
H=F
tcaldate=S
Example of command for calibration coefficients:
dc,OK,46,<CR><LF>
C0=3.24923E-03,E9,<CR><LF>
C1=1.40598E-04,E5,<CR><LF>
…
docaldate=2013/05/21,09,<CR><LF>
A=2.345,59,<CR><LF>
B=13.66666,B4,<CR><LF>
…
tcaldate=2013/05/21,68,<CR><LF>
6. Data format
HEX data format for commands do , tdo , sdo , and stdo ;
The HEX format is extracted as follows.
- Temperatur
e
HEX=TTTT
TTTT is a value for temperature and four-letter HEX value.
Temperature = (decimal expression for TTTT) / 1000 - 5
If temperature < -5.000, then set 0000 to TTTT.
If temperature > 40.000, then set FFFF to TTTT.
- DO
HEX=DDDD
DDDD is a value for DO and four-letter HEX value.
DO = (decimal expression for DDDD) / 100.
If DO > 425.00, then set FFFF to DDDD.
HEX data format for commands tdon , tdona , stdon , and stdona ;
The HEX format is extracted as follows.
- Temperatur
e
HEX is between 0000 and FFFF.
Fixed length of 4
- DO
HEX is between 0000 and FFFF.
Fixed length of 4
- LED accumulated tim
e
HEX is between 00000000 and FFFFFFFF.
Fixed length of 8
The LED accumulated time is expressed by a factor of 10 ms, e.g., the output value “5” means 50 ms.
In order to calculate the LED accumulated time in seconds,
LED accumulated time (s) = (decimal expression for LED accumulated time HEX) / 100
“F” data format
“F” is a real number.
“F” is variable-length and 16 at maximum.
Temperature C coefficient. “F” is a real number.
Temperature B coefficient. “F” is a real number.
Temperature A coefficient. “F” is a real number.
Date of calibration for temperature sensor. The format of “S” is “YYYY/MM/DD”.
Temperature H coefficient. “F” is a real number.
Temperature G coefficient. “F” is a real number.
Temperature D coefficient. “F” is a real number.
d3 coefficient (Uchida et al. , 2013). “F” is a real number.
d4 coefficient (Uchida et al. , 2013). “F” is a real number.
Cp coefficient (Uchida et al., 2013). “F” is a real number.
e0 coefficient (Uchida et al. , 2013). “F” is a real number.
Film number “S” is an ASCII string [0-9a-zA-Z]. The length of the string is 8.
Date of calibration for DO sensor. The format of “S” is “YYYY/MM/DD”.
Description
C0 coefficient (Uchida et al. , 2013). “F” is a real number.
C1 coefficient (Uchida et al. , 2013). “F” is a real number.
C2 coefficient (Uchida et al. , 2013). “F” is a real number.
d0 coefficient (Uchida et al. , 2013). “F” is a real number.
d1 coefficient (Uchida et al. , 2013). “F” is a real number.
d2 coefficient (Uchida et al. , 2013). “F” is a real number.
Temperature F coefficient. “F” is a real number.
Temperature E coefficient. “F” is a real number.
[Appendix 1] 3 /4