EasyManuals Logo

Julabo SL-12 User Manual

Julabo SL-12
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
Menu functions
54
Optimization instructions for the PID control parameters:
optimum setting
°C
t
The heat-up curve reveals inappropriate control
settings.
Inappropriate settings may produce the following heat-up curves:
Xp too low
°C
t
Tv/Tn too low
°C
t
Xp too high or Tv too high
°C
t
Tv/Tn too high or Xp too high
°C
t

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Julabo SL-12 and is the answer not in the manual?

Julabo SL-12 Specifications

General IconGeneral
BrandJulabo
ModelSL-12
CategoryController
LanguageEnglish

Related product manuals