3. Maximum optical output power of the probe: 1.2 mw for infrared light (905 nm).
4.
Bluetooth module: 4.1 (single-mode)
5. Manufacturing date: see the label
6. Normal working condition
5C to 40C (41F to 104F)
15% to 80%, non-condensing
7. Default values and conditions of alert
Upper limit: 100 Lower limit: 94
Upper limit: 130 Lower limit: 50
When the alert switch is on and the actual measured
value goes beyond the preset alert parameter range, the
Oximeter gives an alert sound.
±2% (70% to 100%)
No requirement (≤ 69%)
Upper limit: 50% to 100%
Lower limit: 50% to 100%
Upper limit: 25 bpm to 250 bpm
Lower limit: 25 bpm to 250 bpm
The greater of ±10% of the preset value
and ±5 bpm
Safety Type
Anti-electric-shock type: internal power supply device
Anti-electric-shock degree: BF-type application part
Running mode: continuous working
Waterproof grade: IP22
Storage and Transportation
Packaged products should be stored in well-ventilated rooms without corrosive
gas and with an ambient temperature of –10C to +50C, a relative humidity
10%- 93% (without condensation), and an atmospheric pressure of 50–106 kPa.
After-sale Service
After-sale service unit: Shenzhen Jumper Medical Equipment Co., Ltd.
Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan,
Shenzhen,Guangdong,China
Tel: +86-755-26696279
Fax: +86-755-26852025
E-mail
:
info@jumper-medical.com
Website: www.jumper-medical.com
Postal Code: 518103
Storage and Transportation
ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES
Guidance and manufacturer’s declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The
customer or the user assure that it is used in such an environment.
Electromagnetic environment - guidance
The device use RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The device is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes
Harmonic emissions
lEC61000-3-2
Voltage fluctuations/
Ficker
emissionslEC61000-
3-3
Guidance and manufacturer’s declaration — electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Electromagnetic
environment-guidance
Electrostatic
discharge
(ESD)
lEC61000-4-
2
Floors should be wood , concrete
or ceramic tile. If floors are
covered with
synthetic material, the relative
humidity should be at least 30%
Electrical
fast
transient/bur
st
IEC
61000-4-4
±2 kV for
power supply
lines
± 1 kV
Input/ output
line
not applicable
(For INTERNALLY POWERED
ME EQUIPMENT)
± 1 kV
Differential
mode voltage
± 2 kV
Common
mode voltage
not applicable
(For INTERNALLY POWERED
ME EQUIPMENT)
not applicable
(For INTERNALLY POWERED
interruptions
and voltage
variations on
power
supply input
lines
IEC
61000-4-11
UT)
for 0.5 cycle
40% UT
(60% dip in
UT)
for 5 cycles
70% UT
(30% dip in
UT)
for 25 cycles
<5% UT
(>95% dip in
UT)
for 5 sec
Power
frequency
(50Hz/60Hz
) magnetic
field IEC
61000-4-8
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of device should assure that it is used in such an environment.
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Blood Pressure Monitor,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey(a) ,should be
less than the compliance level in each
frequency range.(b)
Interference may occur in the vicinity of
equipment marked with the following
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and people.
(a) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
Blood Pressure Monitor is used exceeds the applicable RF compliance level above, the
Blood Pressure Monitor should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or
relocating the Blood Pressure Monitor.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
[Vi] V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended
below, according to the maximum output power of the communications equipment.
Maximum
output
power of
transmitter
(W)
Separation distance according to frequency of transmitter(m)
For transmitters rated at a maximum output power not listed above, There commended
separation distance d in meters(m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts(W) accordable to the transmitter manufacturer.
NOTEI At 80 MHz and 800 MHz. the separation distance for the higher frequency
range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and refection from structures, objects and people.