Page 71 of 93
28
Any alarm
message not
listed in this
table.
The ventilator’s
ongoing checks
have detected an
alarm condition.
Check patient; provide alternate
ventilation. Remove ventilator from use
and contact service.
7.4 Alarm Removal
Alarm Condition will automatically reset if the alarm condition is settled.
Operator can silent the alarm to 2 minutes & still the condition not settled with
alarm sound will generate once again.
Current alarm messages and alarm actions will be reset after 5 seconds of Alarm condition
resets.
WARNING
Contact Service alarm will generate continuous beep sound which cannot be silent.
Check Patient and provide alternate ventilation to patient and Remove the ventilator from use
and contact service.
NOTE
Alarm sound cannot be muted without settling Condition.