Step 6: programming in action
Step 7: programming flash finished!!
don’t switch off the chairs whilst programming the flash.
Sometimes an error code will appear on the screen. The USB connection did not connect with the dongle,
cannot initialize CAN interface or another error message. Depending on the version of Windows, you
either need to restart the PC or Laptop and try again. Sometime disconnecting the dongle and
reconnecting it to the PC or laptop will also solve the issue. Just repeat the operation until it connects.
4. Uploading R-net Default files
Upload the R-net OEM files to the wheelchair using R-net OEM software and dongle.
Therefore we now need to open the CxSM developer software and open and write the R-net file once more.
Now all R-net parameters are programmed.
before you open CxSM developer software, first close the OEM R-net software. The dongle can only
handle one program software at a time.
By uploading this file, not all parameters are properly set!