Pull out short form
instructions
Note: For quick alignment reference,
please refer to the short form
instructions located at the rear of the
instrument and accessed by using the
pull out tab on the back left hand side.
10. Measurement procedures
10.3 Taking the reading
9. When two consecutive readings are within 1mmHg an audible
sound will be heard indicating that sufficient readings may have
been taken.
10. If a reading is recorded as a non event or bad event,
a bad event long high pitch tone will be heard.
11. The first reading will be the measured value; successive readings
will display the running average IOP. Outlying or spurious
readings will automatically be excluded from the calculation.
12. At any time the review button will allow you to view the
individual readings.
13. If the unit does not fire, repeat step 3-7.
To measure the other eye, press the OD/OS button on the hand
unit and repeat processes 3 - 7.