3
4
5
6
2
1
10. Measurement procedures
10.1 Preparing the device
1. Plug in the power supply cord to the instrument. The power
socket is located at the rear of the instrument.
2. Turn the instrument on using the on/off push switch located on
the front of the instrument.
3. Lift the handpiece from the cradle.
4. Remove the red protective dust cap from the puff tube.
5. When the hand unit is removed from the holster the two green LED’s on
the front illuminate, and the pump starts, the intelliPuff will perform a
system check, when complete the display will read ‘OK’, refer to section
13 for a full list of display data.
6. Before using the Pulsair intelliPuff press the Demo button to dispel any
minute particles of dust or moisture which may have settled whilst the
Pulsair intelliPuff was not in use.