2260B Series User Manual 
44 
Before setting the OVP or OCP level: 
  Ensure the load is not connected. 
  Ensure the output is set to off. 
1.  Press the OVP/OCP key. The 
OVP/OCP key lights up. 
2.  The OVP setting will be displayed on the top 
and the OCP setting (or OFF) will be displayed 
on the bottom. 
3.  Use the Voltage knob to set the 
OVP level. 
10%~110% of rated 
output voltage. 
4.  Use the Current knob to set the 
OCP level. 
10%~110% of rated 
output current, OFF. 
5.  Press OVP/OCP again to exit. The 
OVP/OCP indicator will turn off. 
Set F-95 (Power switch trip) to 1 (to 
disable the power switch trip) or to 
0 (to enable the power switch trip) 
and save.  
1 (Disable) or 0 (Enable)