OPERATION 
49 
1 = CC High Speed Priority 
3 = CC Slew Rate Priority 
5.  Press the Voltage knob to save the 
configuration setting. ConF will be 
displayed when successful. 
 
6.  If CC Slew Rate Priority was chosen as the 
operating mode, set F-06 (Rising Current Slew 
Rate) and F-07 (Falling Current Slew Rate) and 
save. 
0.01A/s~72.00A/s (2260B-30-36) 
0.01A/s~144.0A/s (2260B-30-72) 
0.01A/s~27.00A/s (2260B-80-13) 
0.01A/s~54.00A/s (2260B-80-27) 
7.  Press the Function key again to exit 
the configuration settings. The 
function key light will turn off. 
8.  Use the Voltage knob to set the 
voltage limit (crossover point).  
Push in the knob to highlight a 
digit. Turn the knob to the desired 
digit to adjust. Rotate the knob to 
the desired voltage. 
9.  Use the Current knob to set the 
current.  
Push in the knob to highlight a 
digit. Turn the knob to the desired 
digit to adjust. Rotate the knob to 
the desired current.