Remote monitoring 
  Voltage monitoring (VMON) and current 
monitoring (IMON) are only supported on 
the master unit. 
  The VMON voltage represents the total 
voltage of the all the serialized units. 
Remote Sense 
  Please see the remote sense chapter for 
details, page 53. 
External Voltage and Resistance Control  
  Voltage/Resistance controlled remote 
control can only be used with the master 
unit. 
Slew Rate 
  The voltage slew rate is actually double that 
of the setting slew rate. I.e., A voltage slew 
rate setting of 60.00V/s is actually 120V/s 
when in series. 
Internal Resistance 
  Only the master unit’s internal resistance 
settings apply.   
  The internal resistance is actually twice that 
of the setting value. 
Bleeder Control 
  The Master unit is used to control the bleeder 
settings. The bleeder resistor is always 
turned on for the slave unit in series mode.