Cycle Step Start: Determines the starting step for the Cycle 
Repeat field. Range: 1~999. The Cycle Step Start field must be less 
than the Cycle Step End field. 
Cycle Step End: Determines the ending step for the Cycle Repeat 
field. Range: 2~1000. The Cycle Step End field must be greater 
than the Cycle Step Start field. 
Line 2: 
Step Header: This field contains the “Step” header text. Required. 
Do not remove from this location. 
Time(sec) Header: This field contains the “Time(sec)” header text. 
Required. Do not remove or alter from this location. 
Voltage (V) Header: This field contains the “Voltage (V)” header 
text. Required. Do not remove or alter from this location. 
Current (A) Header: This field contains the “Current (A)” header 
text. Required. Do not remove or alter from this location. 
Line 3+: 
Step Number: The Step Number field numbers each step in the 
test function. It has a range of 1~1000. All steps must be 
numbered in numerical order. No step number can be “skipped” 
between the first and last step. 
Time(sec): Designates the duration time of the corresponding 
step in seconds. It has a range of 0.05 seconds to 20 days. It has a 
step resolution of 0.05 seconds (50ms). A setting of 0 will skip the 
step. 
Voltage (V): Designates the voltage of the corresponding steps in 
volts. MIN or MAX can be used to set the voltage to minimum or 
maximum, respectively. 
Current (A): Designates the current of the corresponding steps in 
amps. MIN or MAX can be used to set the current to minimum or 
maximum, respectively.