8-12 Status Structure
Operation event status
The used bits of the operation event register (shown in Figure 8-5) are described as follows:
Bit
B1, voltage protection channel #1 (VPT1) — Set bit indicates that the battery
channel (#1) is in voltage protection mode. In this mode, the output has been turned off
and the front panel displays “VPT”. (Battery channel only — for the charger channel, see
Bit 2.)
Bit B2, v
oltage protection channel #2 (VPT2) — Set bit indicates that the charger chan-
nel (#2) is in voltage protection mode. In this mode, the output has been turned off and
the
front panel displays “VPT”. (Charger channel only — for the battery channel, see Bit
1.)
Bit B3,
current limit #1 (CL1) — Set bit indicates that the battery channel’s (#1) output
is in current limit. This bit clears when the instrument is no longer in current limit.
(Battery channel only — for the charger channel, see Bit 7.)
Bit
B4, current limit tripped #1 (CLT1) — Set bit indicates that the battery channel’s
(#1) output has turned off due to a current limit trip condition. This bit clears when the
output is turned back on. (Battery channel only — for the charger channel, see Bit 8.)
Bit B5,
heat sink shutdown (HSS) — This bit indicates that the output has turned off
due to the output stage heat sink overheating.
Bit B6,
power supply shutdown (PSS) — This bit indicates that the output has turned
off due to the main AC/DC power supply heat sink overheating.
Bit B7, current limit #2 (CL2) — Se
t bit indicates that the charger channel’s (#2) output
is in current limit. This bit clears when the instrument is no longer in current limit.
(Charger channel only—for the battery channel, see Bit 3.)
Bit B8,
current limit tripped #2 (CLT2) — Set bit indicates that the charger channel’s
(#2) output has turned off due to a current limit trip condition. This bit clears when the
output is turned back on (charger channel only — for the battery channel, see Bit 4).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2306 and is the answer not in the manual?

Keithley 2306 Specifications

General IconGeneral
BrandKeithley
Model2306
CategoryPower Supply
LanguageEnglish

Related product manuals