Signal Oriented Measurement Commands 10-5
When the MEASure? command is sent, the specified function is selected and then the
READ? is executed. When the MEASure:ARRay? command is sent, the specified function is
selected and the READ:ARRay? command is executed. See READ? and READ:ARRay? for
details.
If a function is not specified, the measurement(s) will be
performed on the active channel’s
function that is presently selected.
NOTE Ther
e are no AVERage commands for long integration measurements. The array size
for long integration readings is fixed at one. Therefore, MEASure:LINTegration? and
MEASure:ARRay:LINTegration? are basically the same.
BOTHTRG Trigger channel 1 then channel 2
BOTHFETCH? Get channel 1 reading then channel 2 read-
ing
BOTHREAD? Trigger both channels then get both read-
ings
When sending either the BOTHTRG or the BOTHREAD? command, make note that the
command is ap
plied to channel 1 (battery channel) first and then to channel 2 (charger channel).
This allows both channels' triggers to be controlled with a single bus command.
The BOTHTRG, BOTHFETCH?, and BOTHREAD? commands work with the
Mo
dels 2306, 2306-VS, and 2306-PJ, but do not work with the
Models 2302 or 2302-PJ (i.e.,
the commands do not work with single channel models).