14-10 Calibration
6. Using the edit keys, adjust the display to agree with the new DMM voltage reading, then 
press ENTER. The unit will prompt for Channel #2 DVM full-scale calibration:
CAL DVM CHAN 2
ALL READY TO DO?
7. Press ENTER to complete Channel #2 DVM full-scale calibration.
8. Connect the digital multimeter volts input and characterized 4Ω resistor 
to the 
Model 2306 OUTPUT #2 SOURCE terminals using the same general connections 
sho
wn in Figure 14-2. Be sure to observe proper polarity (SOURCE + to DMM INPUT 
HI; SOURCE - to INPUT LO).
9. Be sure the digital multimeter DC 
volts function and auto-ranging are still selected.
10. At this point, the unit will prompt for 5A Channel #2 full-scale calibration output:
CAL 5 A CHAN 2
SET 1.90000 A
11. Using the edit keys, adjust the set value to 1.9000A, then press ENTER. The unit will 
prompt you for the DMM reading, which calibrates the Channel #2 5A current limit:
CAL 5 A CHAN 2
READ1 1.90000 A
12. Note the DMM voltage reading, then calculate the current from that reading and the 
actual 4Ω resistance value: I = V/R. Adjust the Model 2306 current display value to 
agree with the calculated current value, and press ENTER.
13. The Model 2306 will prompt for another DMM reading, which is used for Channel #2 
5A measurement calibration:
CAL 5 A CHAN 2
READ2 1.90000 A
14. Again, calculate the current from the new DMM reading and 4Ω resistor value. Adjust 
the 2306 current display reading to agree with the new current, then press ENTER.
15. Disconnect the 4Ω 
resistor, then connect the 3kΩ resistor in its place. (Use the 
connections in Figure 14-3 except make connections to the OUTPUT #2 terminals.)
16. Make sure the DMM DC v
olts function and auto-ranging are still selected.
17. At this point, the unit will prompt to output approximately 5mA for Channel #2 5mA 
range full-scale calibration:
CAL 5 mA CHAN 2
ALL READY TO DO?
18. Press ENTER to output approximately 5mA. The unit will then prompt you for the 
DMM reading:
CAL 5 mA CHAN 2
READ1 4.50000 mA
19. Note the DMM voltage reading, then calculate the current from that voltage reading and 
actual 3kΩ resistance value. Adjust the Model 2306 current display value to agree with 
that value and press ENTER.