C-4 Calibration Reference
:COUNT?
(:CALibration:PROTected:COUNt?)
Purpose To request the number of times the Model 2306 has been calibrated.
Format :cal:prot:count?
Response Number of times calibrated.
Description The :COUNT? query may be used to determine the total number of times the
Model 2306 has been calibrated. The calibration count will also be displayed
during the front panel calibration procedure.
Example :CAL:PROT:COUNT? Request calibration count.
:DATE
(:CALibration:PROTected:DATE)
Purpose To program the calibration date.
Format :cal:prot:date <yyyy>, <mm>, <dd>
Parameters <yyyy> = 1998 to 2097
<mm> = 1 to 12
<dd> = 1 to 31
Query :cal:prot:date?
Response <yyyy> , <mm> , <dd>
Description The :DATE command allows you to store the calibration date in instrument
EEROM for future reference. You can read back the date from the instrument
by using the :DATE? query. The calibration date will also be displayed
during the front panel calibration procedure.
The year, month, and day parameters must be delimited by commas.
Example :CAL:PROT:DATE 2003,11,20 Send cal date (11/20/2003).
:INIT
(:CALibration:PROTected:INIT)
Purpose To initiate calibration.
Format :cal:prot:init
Description The :INIT command initiates the calibration process and must be sent before
all other commands except :CODE.