SECTION2
Operation
NOTE:
The
Q
commands
to create a
sweep
list
(QO
through
Q5)
are
explained
in
paragraph
2.19.3.
HP
BASIC 4.0
Programming
Example:
The
following
program
will
append
a fixed level
sweep
to a staircase sweep.
10 REMOTE716
20
OUTPUT 716; "F0,1X" Source V
Measure
I
(sweep)
30
OUTPUT 716;
"BO,O,X"
Set bias level
to
OV
40
OUTPUT716;
Create
lin
stair
sweep
"Q1,1,10,1,0,500X"
list
50
OUTPUT716;
Append
a fixed level
"Q6,2,0,300,5X"
sweep
60
END
Lines 10
through
40 create a
sweep
list for
the
linear
stair-
case
sweep
waveform.
The
staircase will
start
at
1 V
and
step
in
1V increments
up
to 10V.
The
programmed
dwell
on
each
step
is
1/2
second. Line 50
will
append
a fixed
level (2V, 5 points,
and
300msec delay)
sweep
to
the
stair-
case sweep.
Complete
information
on
the
IEEE-488
bus
and
these
commands
can
be
found
in
Section 3
of
this
manual.
2.21
MODIFY SWEEP
The
modify
feature
permits
source
level
and
delay
time
changes
to
be
made
for
any
measurement
point
in
the
currently defined sweep. This feature allows
custom
shaped
waveforms
to
be
fabricated.
An
example
of
a
modified
sweep
is
shown
in
Figure
2-45.
The
linear staircase
sweep
waveform
is
modified
as
shown
to
demonstrate
the
versatility
of
the
modify
fea-
2-78
+
+
ture.
The
first
four
measure
points
of
the
staircase
are
un-
changed.
The
fifth
point
is
changed
to
the
same
level as
the
fourth
point
(4V).
The
result
is a staircase
step
that
has
more
than
one
measure
point. This allows
multiple
meas-
urements
to
be
performed
on
a single
step
of
a staircase
waveform.
The
rest
of
the
sweep
(points 6, 7
and
8)
is
modified
to
step
down
to
the
bias level
using
a
longer
dwell
time
on
the
steps.
General
Modify
Rules:
1. Enabling MODIFY
displays
the
source
value
of
the
first
sweep
point.
Any
sweep
point
can
be
displayed
with
the
rotary
knob.
SELECT~
and
<IIIII
SELECT tog-
gles
the
display
between
the
source
and
delay
pa-
rameter
for
the
displayed
sweep
point.
2. A
source
value
is
modified
as follows:
A.
Using
the
rotary
knob,
display
the
sweep
point
to
be
modified.
B.
Using
the
keypad,
key
in
the
new
source
value
and
press
ENTER.
C.
Using
the
SELECT
buttons,
display
the
desired
source
range
and
press
ENTER.
3. A
delay
period
is
modified
as follows:
A.
Using
the
rotary
knob,
display
the
sweep
point
to
be
modified.
B.
Using
the
keypad,
key
in
the
new
delay
period
and
press
ENTER.
4. Exit
the
modify
menu
by
again
pressing
MODIFY.
In
many
test situations,
only
minor
modifications to a
sweep
waveform
that
is
already
being
used
will
be
re-
quired.
Complex
waveform
fabrication
will
require
a lit-
tle
more
planning.
The
following
hints
may
be
helpful:
1. Sketch
the
waveform
that
you
wish
to
fabricate. Indi-
cate
source
levels,
measure
points
and
delay
times.
2. Select
and
define (CREATE) a
standard
sweep
waveform
that
will
be
the
easiest
to
modify
into
your
custom
waveform.
3.
Consider
the
value
of
appending
one
or
more
waveforms
to
the
one
you
just
defined
(see
Append
Sweep;
paragraph
2.20).
4.
With
a
good
fundamental
waveform
defined,
go
ahead
and
modify
it
as
follows.