Maintenance and Technical Manual Chapter 1
Interceptor Biological Safety Cabinet Class II A2
October 2019
1.1
CHAPTER 1
Technical Specications for Interceptor
®
Product Number
Width of
Cabinet
Height of
Cabinet
Depth of
Cabinet
Fan
Max
HP
Fan & Light
Max Current/
Power
Outlet
Max Amps
/HZ
Exhaust
CFM
INT-1100A2/INT-1100A2-1
120 VAC
1100mm
43.3"
1610mm
63.4"
813mm
32"
0.9
10.6 Amps
1275 VA
9 Amps
60hz
272
INT-1100A2-230/ INT-1100A2-1-230
230 VAC
5.2 Amps
1275 VA
4.5 Amps
50/60hz
INT-1400A2/INT-1400A2-1
120 VAC
1400mm
55.1"
1610mm
63.4"
813mm
32"
0.9
10.6 Amps
1275 VA
9 Amps
60hz
358
INT-1400A2-230/ INT-1400A2-1-230
230 VAC
5.2 Amps
1275 VA
4.5 Amps
50/60hz
INT-1700A2/INT-1700A2-1
120 VAC
1700mm
66.9"
1610mm
63.4"
813mm
32"
0.9
10.6 Amps
1275 VA
9 Amps
60hz
444
INT-1700A2-230/ INT-1700A2-1-230
230 VAC
5.2 Amps
1275 VA
4.5 Amps
50/60hz
INT-2000A2/INT-2000A2-1
120 VAC
2000mm
78.7"
1610mm
63.4"
813mm
32"
0.9
10.6 Amps
1275 VA
9 Amps
60hz
530
INT-2000A2-230/ INT-2000A2-1-230
230 VAC
5.2 Amps
1275 VA
4.5 Amps
50/60hz
Table 1.1: Product Descriptions
NOTES:
All power values measured at 10 inch operating sash height.
The powered base stand, if ordered, will take an additional 8.6 amps at 120 VAC; 60 Hz or 4.3 amps at 230 VAC;
50 Hz.
The electrical outlets inside the Interceptor
®
are grounded. This is particularly important since all internal surfaces
are stainless steel and conduct electricity. Under NO CIRCUMSTANCES use ungrounded plugs in these outlets.
It is not recommended to use in excess of 1500 watts of power. Exterior power plugs must not be removed until
unit fan and lights are turned o󰀨. The unit is to be disconnected from the main voltage by unplugging both plugs to
remove power. For electrically-powered base stands, power for disconnect is also accomplished by plug removal
from a waste-high or lower plug outlet.
The Interceptor
®
Cabinet has one power cord. If the Power Base is employed, a total of two power cords are
used. When positioning the BSC, always connect BSC and stand power plugs in waste-high or oor-positioned
outlets to facilitate disconnection in an emergency. Never block these outlets.
If UV option is on your BSC, be sure safety overrides are never immobilized! UV lamp should NEVER be on while
sash is open.
continued on page 1.2