3 Turn on the menu ( Shift < ) and then use < or > to select
F: CAL MENU.
4 Use
to move down to the “commands” level and select
2. CALIBRATE.
5 Use
to move down to the “parameters” level and then set the
calibration value for the present input value.
6 Execute the command by pressing
Auto/Man .
7 Perform the appropriate Gain Verification Test to check the
calibration results.
8 Repeat steps 1 through 7 for each gain verification test point
shown in the tables. Each range in the gain adjustment procedure
takes less than 20 seconds to complete.
ENTER
4
Chapter 4 Calibration Procedures
Gain Adjustment
81