2 Verification and Adjustments
76 Keysight E36200 Series Service Guide
Table 2-12 Test record form - Keysight E36232A
E36232A Report number _______________ Date _______________
Description DUT setting Lower limit Result Upper limit
Constant voltage tests
Voltage programming
Zero voltage output (V
0
)
0 V, 10 A –0.008 V ____________ 0.008 V
Maximum voltage output (V
max
)
60 V 10 A 59.974 V ____________ 60.026 V
Voltage readback
Zero voltage measured over interface 0 V, 10 A
V
0
– 0.008 V
____________
V
0
+ 0.008 V
Maximum voltage measured over interface 60 V 10 A
V
max
– 0.026 V
____________
V
max
+ 0.026 V
CV load regulation (V
load
- V
noload
)
60 V, Max A –8 mV ____________ 8 mV
20 V, Max A –4 mV ____________ 4 mV
CV line regulation (V
lowline
- V
highline
)
60 V, Max A –8 mV ____________ 8 mV
20 V, Max A –4 mV ____________ 4 mV
CV ripple and noise
peak-to-peak
60 V, Max A - ____________ 3.5 mV
20 V, Max A - ____________ 3.5 mV
rms
60 V, Max A - ____________ 0.35 mV
20 V, Max A - ____________ 0.35 mV
Transient response
60 V, Max A - ____________ 50 μs
20 V, Max A - ____________ 50 μs