Installation 2
Series E4360 User’s Guide 25
Bench Installation
CAUTION
Do not block the air intake and exhaust at the sides, or the exhaust at
the rear of the unit. Refer to the outline diagram in Appendix A.
Minimum clearances for bench operation are 2 inches (51 mm) along the
sides and back.
400 Hz Operation
Redundant Ground Requirement
At 400 Hz AC input operation, the leakage current of the unit exceeds
3.5 mA. This requires the installation of a permanent, redundant ground
from the instrument chassis to earth ground. This ensures that ground
will always be connected and that any leakage current will be diverted to
ground. Refer to the Service Guide for installation instructions.
Power Factor
Refer to Appendix A for power factor statistics at 400 Hz operation.