Error Messages Appendix A
Series E4360 Service Guide 67
Device-dependent Errors (continued)
116
Locked out by internal switch setting
Calibration mode is locked out by the calibration switch. Refer to the Service Guide.
117
Calibration Error
An error occurred during calibration. Exit calibration mode and do not save calibration settings.
118
Must be in Fixed mode to calibrate
The instrument must be set to Fixed mode operation before it can be calibrated.
200
Hardware error channel <channel>
A hardware error has occurred on the specified channel.
201
Invalid configuration, empty slots
There is an empty slot between modules. This configuration is not allowed.
202
Selftest Fail
A selftest failure has occurred. See selftest failure list for details.
204
NVRAM checksum error
A checksum error has occurred in the instrument’s nonvolatile random access memory.
205
NVRAM full
The nonvolatile random access memory of the instrument is full.
206
File not found
The internal calibration file or the internal channel attribute file was not found in NVRAM.
207
Cal file version error
The calibration file was written or read using old firmware. Firmware must be updated.
208
Running backup firmware
Backup firmware is running because something is wrong with the primary firmware. Firmware must
be updated.
209
Modules do not match group NV config
A module has been replaced in the mainframe that does not match the characteristics required for
it to be a grouped module.
210
Frame NVRAM error
An error has occurred in the mainframe's nonvolatile random access memory.
302
Option not installed
The option that is programmed by this command is not installed.
303
There is not a valid acquisition to fetch from
There is no valid data in the measurement buffer.
304
Volt and curr in incompatible transient modes
Voltage and current cannot be in Step and List mode at the same time.
305
A triggered value is on a different range
A triggered value is on a different range than the one that is presently set.
306
Too many list points
Too many list points have been specified.
307
List lengths are not equivalent
One or more lists are not the same length.
308
This setting cannot be changed while transient trigger is initiated
Setting cannot be changed while the instrument is waiting for or executing a trigger sequence.
309
Cannot initiate, voltage and current in fixed mode
Cannot initiate the transient generator because either the voltage or the current function is set to
Fixed mode. Transient triggering is not available in Fixed mode.
311
This group of channels has non-identical features
The characteristics of the modules that have been selected to be grouped do not match.
312
The command is disabled when the channel is grouped
The command is disabled when the channel is grouped. Ungroup the channel to use command.