EasyManuals Logo

Keysight Technologies FieldFox A Series User Manual

Keysight Technologies FieldFox A Series
756 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #558 background imageLoading...
Page #558 background image
558 Keysight N9927-90001 User’s Guide
Noise Figure Mode (Option 356 – CPU2 Only)
How to Define a Noise Figure Measurement
T Cold Default When making measurements at different
ambient temperatures, you can set the T
cold
value in
Kelvin (K) manually (User). When set to Default, the T
cold
value is set to the 296.5 K (Default = Default).
Back
to return to previous menu.
DUT
or
Opens a menu of softkeys to define your DUT Type (Amplifier,
Downconverter, Upconverter, or Multi-Stage Converter), frequency
context, and loss compensation. Refer to “How to Use DUT
Measurement Setups” on page 562.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies FieldFox A Series and is the answer not in the manual?

Keysight Technologies FieldFox A Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelFieldFox A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals