174 Keysight FieldFox Handheld Analyzers Service Guide
General Purpose Maintenance and Service Procedures
Firmware Upgrades
8-
Firmware Upgrades
To Check the Current Firmware Revision
—Press System(7).
Press Service Diagnostics.
—Press System Information.
The current firmware version is indicated on the “Firmware Release” line.
For information on upgrading FieldFox firmware, go to
www.keysight.com/find/fieldfoxsupport.
If the firmware upgrade process is interrupted, memory contents in the
instrument may become corrupted. If a problem is suspected, consider doing a
fresh installation of the firmware. Follow the process under “Install Firmware
from SD Card” near the end of Chapter 7.
See the “Firmware Related Information” section of Chapter 5 for details on
what happens during the firmware update process.
Miscellaneous Service and Maintenance
Cleaning the Display
Materials: Use a soft, lint-free cloth. Use a solution with 99.5% IPA (isopropyl
alcohol) or a high quality window cleaning solution.
Technique: Do not rub back and forth with a polishing motion. Use a single
swipe from the soiled location to the nearest edge of the display.
Battery Maintenance
See the “Working with the Lithium-Ion Battery” section of the FieldFox User’s
Guide for information on the removable battery. In particular, see “Built-In
Battery Gauge” and “Reconditioning the Battery”.