Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 203
RF Assembly
A3 RF Assembly Troubleshooting
Figure 10-3 Synthesizer Multiplier / Divider Section Output – Vector
4. Tune the instrument to a frequency that exhibits the power level issue.
5. Connect a spectrum analyzer to A3 J2004.
6. Configure the spectrum analyzer with the following settings:
Center Frequency = Frequency of signal source
Span = 10 MHz
Reference Level = +10 dBm
7. Turn the signal path on by pressing I/Q, I/Q On.
8. The power level of the signal at A3 J2004 should be at a fixed level of 0
dBm, +/-6dB.
If the power level measured is not 0 dBm, +/-6 dBm, replace the A3
RF assembly since there are no adjustments for this power level.
If the power level measured is 0 dBm, +/-6 dB, the power level at
the output of the Synthesizer Multiplier / Divider section is correct.
The issue is somewhere farther down the signal path.