338 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Assembly Replacement
Inner Top Cover
Inner Top Cover
Tools Required
T-10 driver
Removal Procedure
Refer to Figure 15-2 for this procedure.
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to “Outer Cover”
on page 336.
3. Place the signal generator flat with the front-panel facing you.
4. Using the T-10 driver, remove the eleven flat-top screws (1) from the
inner top cover (2).
5. Using the T-10 driver, remove the screw (3) from the rear panel.
6. Remove the inner top cover.
Replacement Procedure
1. Reverse the order of the removal procedures.
2. Recommended torque for all screws is 9 in–lbs.