EasyManua.ls Logo

Keysight 34401A - DC Characteristics

Keysight 34401A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
_
|
Chapitre 1 : Pages 12-13 - DC Characteristics and Measuring/Operating Characteristics
Ajoutez une note de bas de page [12] sous le paramètre “Function” “ Diode Test” (page 12) et le texte suivant
à la page 13 :
“[12] Accuracy specifications are for the voltage measured at the input terminals only. 1mA test current is
typical. Variation in the current source will create some variation in the voltage drop across a diode junction.”
Chapitre 1 : Page 17 - Frequency/Period Measuring/Operating Characteristics
Changez : [4] Input > 100 mV. For 10 mV input, multiply % of reading error x10.
En : [4] Input > 100 mV. For 10 mV to 100 mV inputs multiply % of reading error x10.
Chapitre 1 : Page 18 – General Information
Changez : Warranty: 3 years standard
En : Warranty: 1 year standard
Chapitre 1 : Page 19 – Product Dimensions
1. Changez les dimensions dans le diagramme supérieur comme suit :
2. Recherchez et modifiez la dimension du panneau avant de ‘42.58’ (mm) à ‘67.53’ (mm) dans la partie
supérieure gauche du diagramme inférieur.
Chapitre 6 : Page 112 - Types of Service Available
Changez : “If your multimeter fails within three years of original purchase…”
En : “If your multimeter fails within one year of original purchase…”
Changez : “If your unit fails after your three year warranty expires …”
En : “If your unit fails after your one year warranty expires …”
Chapitre 6 : Page 113 – Précédant la section Repackaging for Shipment
Changez : “…, will continue the warranty time clock to the end of your three year standard warranty.”
En : “…, will continue the warranty time clock to the end of your one year standard warranty.”
Changez : “If your three year warranty has expired…”
En : “If your one year warranty has expired…”
Chapitre 6 : Page 114 – To Replace the Power-Line Fuse
Supprimez la dernière phrase “To order the 125 mAT slow-blow fuse, use Agilent part number 2110-0894.”