Capitolo 6: Pagina 112 - Types of Service Available
Modificare: “If your multimeter fails within three years of original purchase…”
In: “If your multimeter fails within one year of original purchase…”
Modificare: “If your unit fails after your three year warranty expires …”
In: “If your unit fails after your one year warranty expires …”
Capitolo 6: Pagina 113 - Prima della sezione Repackaging for Shipment
Modificare: “…, will continue the warranty time clock to the end of your three year standard warranty.”
In: “…, will continue the warranty time clock to the end of your one year standard warranty.”
Modificare: “If your three year warranty has expired…”
In: “If your one year warranty has expired…”
Capitolo 6: Pagina 114 – To Replace the Power-Line Fuse
Eliminare l'ultima frase “To order the 125 mAT slow-blow fuse, use Agilent part number 2110-0894.”
Capitolo 7: Pagina 132 - 34401-66501 Main PC Assembly (A1) Replaceable Parts
Modificare: U502 34401-88861 Prog EPROM (1818-5187) 28480 34401-88861
(See above) 1818-5187 IC 2M OTP ROM 150-ns CMOS 34335 AM27C020-155JC
In: U502 34401-88806 Prog EPROM (1818-5680) 28480 34401-88806
(See above) 1818-5680 IC 2M-BIT OTP 120-ns CMOS 04078 M27C2001-12C1E
Modificare: U701 1820-6974 IC-GPIB Controller IEEE-488 1975/78
In: U701 1822-0639 IC-GPIB Controller IEEE-488 1975/78
Capitolo 7: Pagina 134 - Agilent 34401A Mainframe Replaceable Part
Agilent 34401A Mainframe
Modificare: CVR1 34401-84101 Instrument Cover In: 34401-84111
FRM1 34401-80101 Chassis 34401-80121
MP1 03478-88304 Rear Panel / Bezel Assembly 34401-88304
MP2 34401-40201 Front Panel / Bezel Assembly 34401-40231
MP3 34401-43701 Power-on Push Rod 34401-43731
MP4 34401-43702 Front / Rear Push Rod 34401-43732
MP5 34401-45001 Carrying Handle 34401-45021
MP8 34420-49301 Window, Front Panel Display 34420-49311
MP9 34401-81902 Flubber Keys – Front Panel 34401-81932
MP11-MP12 34401-86000 Bumpers (Front/Rear) & Power Mod Cvr 34401-86020
Exchange Unit Rebuilt 34401A 34401-69211
Capitolo 7: Pagina 134 - Agilent 34401A Mainframe Replaceable Parts
Modificare la nota in fondo alla tabella Mainframe Agilent 34401A come segue: “* For serial numbers prior to
3146A31949, use a 3A, 250 Vac fuse (p/n 2110-0780) for 220 or 240 Vac operation; for all other serial
numbers, use a 250 mAT fuse (p/n 2110-0817) for all line voltages.”
Capitolo 8: Pagina 139 - 34401-66522 Display Assembly Replaceable Parts List
Modificare: 34401-66502 Display Assembly Replaceable Parts List
In: 34401-66522 Display Assembly Replaceable Parts List