These manual updates apply to the 34401A User’s Guides listed below. Updates are also provided for Edition
5 of the 34401A Service Guide (p/n 34401-90013) starting on page two. Page numbers correspond to the
English version of the manual.
• 34401-90413 (User’s Guide – Japanese
)
• 34401-90416 (User’s Guide – Korean)
• 34401-90417 (User’s Guide – Taiwanese)
• 34401-90418 (User’s Guide – German
)
• 34401-90419 (User’s Guide – French)
• 34401-90420 (User’s Guide – Spanish)
• 34401-90421 (User’s Guide – Italia
n)
• 34401-90422 (User’s Guide – Chinese)
• 34401-90438 (User’s Guide – Russian
)
UUpdates – 34401A User’s Guide
Product Warranty
The standard warranty period for the 34401A Multimeter is one-year. All references to a three-year warranty
within the product manuals must be changed to one year.
Chapter 3: Page 76 - The Wait-for-Trigger State
Add the word “external” to the second sentence in the boxed note at the end of The Wait-for-Trigger State
section as follows: Any external triggers that occur during this set-up time are ignored.
Chapter 3: Page 100 - To Replace the Power-Line Fuse
Delete the last sentence “To replace a 125 mAT fuse, order Agilent part number 2110-0894.”
Chapter 4: Page 123 - [SENSe:DETector:BANDwidth?] [MINimum | MAXimum]
[SENSe:DETector:BANDwidth?] [MINimum | MAXimum]
Query the ac filter. Returns “+3.000000E+00”, “+2.000000E+01”, or “+2.000000E+02”.
Chapter 4: Page 148 - RS-232 Configuration Overview
Parity and Data Bits: None/8 data bits
Even/7 data bits (factory setting), or
Odd/7 data bits
Chapter 5: Page 178 - Execution Error 550
Command not allowed in local
The multimeter received a READ? or MEASure? command while in the local mode. During RS-232 operation,
you should always execute the SYSTem:REMote command before sending other commands over the interface.