Frequency Mode AUTO for frequency, frequency ratio, or average period, PRF
or PRI measurements. Unchanged otherwise.
<function> Gate Source Depends on measurement function. See the CONFigure
commands for the different measurement functions for
details.
Gate Start Source IMMediate
Gate Start Delay TIME, 0 sec
Gate Stop Source IMMediate
Gate Stop Holdoff TIME, 0 sec
Gate Start and Stop Slope NEGative
External Gate Source Gate In/Out BNC
Burst Gate Auto Enabled
Burst Gate Delay 0.0
Burst Gate Time 1 uS
Narrow Pulse Mode Disabled
Samples per Trigger
1 sample, except for the CONFigure:ARRay:TSTamp command,
which sets this value to the <count> parameter for this command.
Trigger Count 1 trigger
Trigger Delay 0 sec
Trigger Source IMMediate
Trigger Slope NEGative
Timestamp Rate 1.0E6
CALCulate subsystems Disabled, including the individual calculate functions. The other parameters
are unchanged.
Reference Oscillator Unchanged
STATus, DATA, and FORMat sub-
systems
Unchanged
Auto-level Frequency Unchanged
Measurement Timeout Unchanged
Using the CONFigure Command
The following program segment shows how to use the CONFigure command with the READ? command to
make an externally-triggered measurement. The CONFigure command configures the instrument for fre-
quency measurements. Note that the CONFigure command does not place the instrument in the "wait-
for-trigger" state.
Keysight 53220A/53230A Programmer's Reference 113
CONFigure Subsystem Introduction