SYSTem:COMMunicate:LAN:TELNet:WMESsage
Syntax
SYSTem:COMMunicate:LAN:TELNet:WMESsage "<string>"
SYSTem:COMMunicate:LAN:TELNet:WMESsage?
Description
This command sets the welcome message seen when using a Telnet session to communicate with the
instrument.
Parameters
Name Type Range of Values Default Value
<string> Quoted String A string of up to 63 characters. This is a required parameter
Remarks
l The instrument uses LAN port 5024 for SCPI Telnet sessions (but 5025 for SCPI Socket sessions).
l The welcome message is stored in non-volatile memory, and does not change when power has been
off, after a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet command).
l The welcome message is set to "Welcome to Keysight's 53220A Universal Counter/Timer" (Keysight
53220A) or "Welcome to Keysight's 53230A Universal Counter/Timer" (Keysight 53230A) when the
instrument is shipped from the factory or after a SYSTem:SECurity:IMMediate command.
Return Format
The query reads the welcome message and returns an ASCII string enclosed in double quotes.
Examples
The following example defines the welcome message.
SYST:COMM:LAN:TELN:WMES "Welcome to the 53220A Telnet Session"
See Also
SYSTem:COMMunicate:LAN:TELNet:PROMpt
Keysight 53220A/53230A Programmer's Reference 421